Friday 26 February 2021

Lupine Publishers | On Mammography and Hormone Replacement Therapy

 Lupine Publishers | Journal of Gynaecology


Introduction

About 250 years ago, French philosopher Francois-Marie Arouet Voltaire, wrote:

“Doctors prescribe medicine of which they know little, to cure diseases of which they know less, in human beings, of which they know nothing”.

Has this state of affairs changed very much since then? I believe so. However, many would argue “hardly”, aside from new technological advances and certain heroic surgical procedures. Our model continues to be population-based (also known as “mass medicalization”) rather than patient-centric. Further, notwithstanding the current dogma on “evidence-based” medicine, meaning that the health care provided is based on solid scientific evidence of utility, a large proportion of tests and prescriptions used frequently have little or no such supportive evidence. Another flaw of today’s “evidence-based” medicine is what has been termed “eminence-based” medicine wherein experts make recommendations or “guidelines” for a large proportion of decisions for which no or minimal data exists. These guidelines have a pronounced impact, as they are believed to represent the standard of care, even though they are based on opinion with a paucity of facts. Actually, even the prestigious U.S. Institute of Medicine concluded that “any valid evidence supports “well below half” of the practice of medicine (!)”. Examples abound such as, statins and statin combinations, prostate specific antigen tests, mammography, hormone replacement therapy, etc., I will discuss here mammography and hormone replacement therapy, which are of particular importance in women’s healthcare.

Mammography

Mammography for breast cancer screening for women parallels the prostate specific antigen (PSA) test in men. This imaging test uses low-energy X-radiation to examine the human (female but also male) breast. It is used both as a screening and a diagnostic test. The goal of any screening procedure is to examine a large population of patients to find that small number most likely to have a serious condition. These patients are then referred for further, usually more invasive, testing. Thus a screening exam is not intended to be definitive, rather to have sufficient sensitivity to detect a useful proportion of cancers. The cost of higher sensitivity is a larger number of results that would be regarded as suspicious in patients without disease. In mammography, the goal is the early detection of breast cancer through the detection of characteristic masses and/or micro-calcifications. Its use as a screening tool for the detection of early breast cancer in otherwise healthy women without symptoms is controversial. Like all X-rays, mammograms use doses of ionizing radiation (lower than those employed in bone radiography) to create images that are subsequently analyzed for any abnormal findings.

A. Adjunct procedures to mammography are:

a. Ultrasound: For further evaluation of masses, including palpable masses not seen on mammograms;

b. Ductography (not generally used): For further evaluation of questionable findings as well as for screening pre-surgical evaluation in patients with known breast cancer to detect any additional lesions that might change the surgical approach, for instance from breast-conserving lumpectomy to mastectomy;

c. Magnetic resonance mammography: For greater spatial resolution of mammographic tissue imaging;

d. Positron emission mammography; and

e. New procedures, including breast tomosynthesis.

B. Currently recommended guidelines for having mammography screening tests for the average woman are:

a. U.S. Preventive Services Task Force (2009): Screening of women aged between 40 and 49 should not be routine but based on individual’s risk factors and values (because the benefits of screenings do not outweigh the risks). Every two years between the ages of 50 and 74;

b. American Cancer Society, American College of Radiology, American Congress of Obstetricians and Gynecologists: Annually beginning at age 40;

c. National Cancer Institute: Every one to two years for women ages 40 to 49;

d. American College of Physicians: Individualized screening plans as opposed to wholesale biannual screening of women aged 40 to 49;

e. Canadian Task Force on Preventive Health Care (2012): Every 2-3 years between the ages of 50 and 69. It found that for women aged 50-69, screening 720 women once every 2-3 years for 11 years would prevent 1 death from breast cancer. For women age 40-49, 2100 women would need to be screened at the same frequency and period to prevent 1 death from breast cancer; and

f. European Cancer Observatory (2011): Every 2-3 years between the ages of 50 and 69.

The reports from the above task forces note that the risks of more frequent mammograms include a small but significant increase in breast cancer induced by radiation, a risk that is greater for younger women. On the other hand, the Cochrane Collaboration (2011) analysis of screening further concluded that: “Mammograms reduce mortality from breast cancer by an absolute amount of 0.05% or a relative amount of 15%, but also result in unnecessary surgery and anxiety such that it is not clear whether mammography screening does more good than harm... and that universal screening may not be reasonable”. It also states that “the best quality evidence does not demonstrate a reduction in mortality generally or a reduction in mortality from all types of cancer from screening mammography”. In addition, the Nordic Cochrane Collection (2012) states that “advances in diagnosis and treatment make mammography screening no longer effective today in decreasing deaths in breast cancer, and therefore no longer recommend routine screening for healthy women at any age as the risks might outweigh the benefits... and warns of misleading information on the internet”. Further, their analysis showed that “one in 2,000 women will have her life prolonged by 10 years of screening, however, another 10 healthy women will undergo unnecessary breast cancer treatment. Additionally, 200 women will suffer from significant psychological stress due to false positive results”.

Repeated mammography starting at age 50 saves about 1.8 lives over 15 years for every 1,000 women screened. This result must be gauged against the negatives of errors in diagnosis, overtreatment and radiation exposure. Also, screening mammography does not reduce death overall, but causes significant harm by inflicting cancer scare and unnecessary surgical interventions. About 7% (more realistically, 10%-15%) of women screened with mammography will be called back (with great distress) for a diagnostic session. However, most of these recalls will result in “false positive” results. For 1,000 recalls, about 60 will have benign growths and 10 will be referred for a biopsy (of which about 3.5 will have a cancer of which about 2 will be a low stage cancer that will be essentially cured after treatment, and 6.5 will not). Mammography may also produce “false negatives” (not seeing the cancer), usually around 10%-30%, and due to

(a) observer error,

(b) cancer hidden by other dense tissue in the breast, and

(c) Cancer overlapping normal tissues.

Furthermore, one form of breast cancer, lobular cancer, has a growth pattern that produces shadows on the mammogram which are indistinguishable from normal breast tissue. A meta-analysis review of programs in countries with organized screening found 52% over-diagnosis. Women whose breast cancer was detected by screening mammography before the appearance of a lump or other symptoms commonly assume that the mammogram “saved their lives”. In practice, the vast majority of these women received no practical benefit from the mammogram. There are four categories of cancers found by mammography:

i. Cancers that are so easily treated that a later detection would have produced the same total cure (that is, the woman would have lived even without mammography);

ii. Cancers so aggressive that even “early” detection is too late (the woman dies despite detection by mammography);

iii. Cancers that would have receded on their own or are so slow-growing that the woman would die of other causes before the cancer produces symptoms (mammography results in overdiagnosis and over-treatment); and

iv. The small number of breast cancers that are detected by screening mammography and whose treatment outcome improves as a result of earlier detection. Clinical trial data suggest that 1 woman per 1,000 healthy women screened over 10 years falls into this category. Screening mammography produces no benefit to any of the remaining 87% to 97% of women.

C. In summary

The guidelines for screening mammography advocated by the several professional associations or/and governmental organizations are conflicting and even confusing. Would it not be helpful for patients if these entities were to agree to a uniform set of guidelines (even though these would still be “guidelines”)? Further, because mass screening as a tool for the detection of early breast cancer in otherwise healthy women without symptoms is controversial, shouldn’t this screening be conducted on an individual basis and only in case of significant risk? Still further, since the radiation sensitivity of the breast in women under age 35 is greater than in older women, should it not be generally imperative that these women be screened only if there is a significant risk of cancer (such as, BrCa positive, very positive family history, palpable mass) and even in these circumstances to employ ultrasound or magnetic resonance for imaging? Also, and likewise, should screening of women aged between 40 and 49 not be routine but based on individual’s risk factors and values (because the benefits of screenings do not outweigh the risks)? Additionally, beyond age 50, should screening not be conducted systematically and only infrequently at appropriate time intervals to be defined? Lastly,based on the important Cochrane Collaboration and the Nordic Cochrane Collection, should not routine screening be discouraged for healthy women of any age as the risks might outweigh the benefits? The above provides more evidence that population medicine (in this case, mass screening) disregards individual variability and promotes considerably more unnecessary medical testing and procedures.

Conclusion

a. Since the guidelines for screening mammography advocated by the several professional associations or/and governmental organizations are conflicting and even confusing, this screening should be conducted on an individual basis and only in case of significant risk.

b. Since the radiation sensitivity of the breast in women under age 35 is possibly greater than in older women, it would be generally imperative that these women be screened only if there is a significant risk of cancer (such as, BrCa positive, very positive family history, palpable mass), and even in these circumstances the screening should employ ultrasound or magnetic resonance for imaging.

c. Screening of women aged between 40 and 49 should not be routine but based on individual’s risk factors and values (because the benefits of screenings do not outweigh the risks).

d. Beyond age 50, screening should not be conducted systematically but only infrequently at appropriate time intervals to be defined.

e. Based on the important Cochrane Collaboration and the Nordic Cochrane Collection, routine screening should be discouraged for healthy women of any age as the risks might outweigh the benefits. Whereas much more could be said about the mammography test, I would now like to proceed to the example of hormone replacement therapy.

Hormone Replacement Therapy

Hormone replacement therapy (HRT) refers to any form of hormone therapy wherein, in the course of medical treatment, the patient receives hormones, either to supplement a lack of naturally occurring hormones or to substitute other hormones for naturally occurring hormones. We are here interested solely in HRT for menopausal women. The idea is that treatment may prevent the discomfort caused by diminished circulating estrogen and progesterone hormones or, in the case of the surgically or prematurely menopausal women, that it may prolong life and may reduce the incidence of dementia. It involves the use of one or more of a group of medications designed to artificially boost hormone levels. The main types of hormones involved are estrogens, progesterone or progestins, and sometimes testosterone. It is often referred to as “treatment” rather than therapy.

Many studies on the effects of HRT have been conducted on rats. Overall, the results of these studies are non-conclusive and more research in this area is needed. Nonetheless, some important results can be gathered:

a. Differing brain regions may respond in a variety of ways to HRT;

b. Timing of the therapy is integral to the chances of success; and

c. How the hormones are administered, either chronically or cyclically, may make an important difference in their effectiveness.

As recently as 2005, women have had a positive and overly optimistic attitude towards HRT. Currently, however, most women do not find HRT to be an effective solution: It is initially helpful but, if used for a long period of time, it loses its effectiveness, and there are times when it is not only ineffective but actually detrimental to people. In the case of menopausal women, HRT has had the following adverse effects:

a. Impaired hearing including decrease in the functionality of many regions of the ear,

b. Reduction of the effectiveness in parts of the central nervous system used for hearing, and

c. Increased chance for cardiovascular disease (particularly in the case of women caregivers who experience more acute stress in their lives).

However, HRT can have beneficial effects:

a. Positive effects on the prefrontal cortex by boosting the working memory,

b. No additional weight gain compared to women who do not use HRT, and

c. Positive effects in their sex life (mainly increasing their sex drive and sexual sensitivity) that can dissipate after receiving HRT for extended periods of time ... but the effects are inconsistent across women. For decades, HRT was widely recommended to women to reduce heart disease.

However, the Women’s Health Initiative (WHI) trial (over 16,000 post-menopausal women) compared the combination (estrogen + progestin) to placebo. The findings included significant increases in breast cancer, heart disease and heart attacks, strokes, and dangerous blood clots. These findings far over-rode the alleged benefit of less colon cancer and fewer hip fractures. The results of the WHI trial were so negative that it was stopped prematurely at 5.6 years (instead of the planned 15 years) of follow-up. New results released in 2011 continue to engender confusion, suggesting disparate outcomes with hormone replacement as a function of what age the treatment was initiated.

Nonetheless, after a slowing period following the announcement of the trial’s negative results, the practice continues under the guise (true perhaps for some women) that HRT (estrogen and/or progestin) palliates the unpleasant “hot flashes” post-menopausal women experience. But, this latter “benefit”, even if true, is not the premise on which HRT was advocated and sold. Of last note, a recent report in the Archives of Internal Medicine revealed that “post-menopausal women who were treated with statin drugs to lower their cholesterol had a nearly 48% increased risk of developing diabetes compared to those who were not given the drug. This is even more critical when we consider that becoming diabetic doubles the risk for Alzheimer’s disease. The combination (HRT + statins) is of serious concern.

Summary

Whereas initially helpful, HRT loses its effectiveness when used for a long period of time, and there are times when it is not only ineffective but actually detrimental. In the case of menopausal women, HRT has multiple adverse effects (impaired hearing including decrease in the functionality of many regions of the ear, reduction of the effectiveness in parts of the central nervous system used for hearing, and increased chance for cardiovascular disease particularly in the case of highly stressed women). However, it can also have beneficial effects (boosting the working memory, no additional weight gain, and positive effects in the sex life) that unfortunately can dissipate after receiving HRT for extended periods of time. Whereas significant increases in breast cancer, heart disease and heart attacks, strokes, and dangerous blood clots were found in a large and important clinical trial, HRT was discontinued for a time period but is witnessing resurgence for alleged other benefits (palliation of hot flashes). Again, even when the surrogate end point is no longer tenable, another surrogate end point is found to justify the continued use of the therapy (a common marketing ploy). What then is the key to the change in this population medicine paradigm? It resides in the 6 rights: “right” doctor, “right” screen test, “right” patient, “right” drug, “right dose”, and “right” cost. This should be tackled and hopefully implemented.

Read More Gynaecology and Women's Health Care Lupine Publishers Articles:
https://lupinepublishers-gynecology.blogspot.com/

Thursday 25 February 2021

Lupine Publishers | Comparative Analysis of Development Program of Concepts and Categories Through Cognitive Perceptive Networks in Students with Autism Spectrum Disorder

 Lupine Publishers | Journal of Psychology


Abstract

Individuals with autism spectrum disorder (ASD) exhibit social- communicative impairments, which cognitive and neuropsychological deficits are evident. This research study presents an experimental study to prove effectiveness of a psychologic program, designed to facilitate conceptual and categorical development through learning of links between concepts. A total of 27 students with ASD participated in study, evaluated over 9 months, throughout three successive measures, distributed in two groups: an experimental group (n= 14) and a control group (n= 13). Results found through ANOVA comparative analysis and independent samples t- test for equality of means show there´re significant differences in the cognitive conceptual development among experimental group participants compared to their peers of control group.

Keywords: Autism; Perception; Cognition; Sensory Integration; Semantic Memory

Introduction

According to International Classification of Mental Disorders of American Psychiatric Association [1], people with autism spectrum disorder (ASD) show a characteristic perceptual processing, which also extends to evaluation of experience, both external and internal, since processes of analysis are conditioned, both the information coming from environment, as the existing contents in permanent memory. Mevel, Fransson and Bölte (2015) affirm that these deficits are due to complex interaction of genetic and environmental factors, which affect cerebral maturation, synaptic function and cortical networks. [2] deepen into an etiological study and conclude that analysis performed with a comparative genomic hybridization matrix (array-CGH), indicates that copy number variants (CNV) are associated with the susceptibility of people with ASD, which is an important specificity of these individuals. Likewise, [3] points out that cognitive processing follows a sequence in which, neuronal system determines strength of signals detected, so process transforms initial visual stimulus into neuronal impulses, which are transformed from retina to the brain in many ways, of which, the most remarkable is the lateral geniculate nucleus. Then, arrive to primary visual cortex, where it occurs analysis process of individual features observed of initial stimulus. From this moment, it´s distributed to several areas, that process the isolated features, which are combined in multiple ways to configure a global representation of initial image of retina. But, there hasn´t yet been conceptual- categorial attribution, thus, still must go to an intermediate and superior perceptual level, where inferior temporal cortex will develop an essential performance, so any deficit on any area of lower temporal cortex can cause deficits to the recognition of perceived object [4] have explored specificity of the activation of facial processing in people with ASD and its results demonstrate a hypoactivation in facial processing system of human faces, but not of animals, possibly due to limitations in ability to represent value of social stimuli [5] say these neurological deficits aren´t limited to one or several domains, but in their interaction, exhibiting limited cerebral axes, related with ability to modify attention and executive function, while indicating better potential in activities planning [6].

Deepen into the possible causes of conceptual integration deficits, which´s explained by the cognitive hypotheses of coherence and conclude that people with ASD during joint attention processes showed reduced temporal-central alpha coherence in right hemisphere compared to their peers in the typical development group. Then, studies about presence of a local perception are corroborated, just as the theory of cognitive central coherence affirms [7]. Likewise, these hypotheses were analyzed by [8], that suggest global integration is reduced, being an important factor of weak central coherence. Whereby, executive system is affected, with great limitations for conceptual integration and categorical hierarchy. In this sense [9], investigate the computational work and suggests these deficits may be result of atypical representations in people with ASD in cortical maps, derived the idiosyncratic perceptual processes observed.

Aims

Adapted various methodological proposals related to people with ASD to respond to these specific needs [10,11]. However, a recurring theme is existence of an empty in programs design to create links between the concepts; therefore, this research´s main aim is to design a specific program to facilitate of conceptual and sensory integration development, through creation of relational links or networks. Thus, basic aims are the following:

A. Design and apply a specific program to facilitate conceptual sensory integration, through creation of perceptivecognitive networks to people with ASD.

B. Evaluate the application of program on the experimental group participants, compared with his peers of control group.

Method

Design: Research design is supported by an experimental model of two groups, 1 Experimental Group (EG), to which a specific program was applied, and 1 Control Group (CG), whose participants followed the regular programs.

Participants: A total of 27 students with ASD participated in study, of three diagnosis levels, of both sexes and of ages between 4-15 years, which have been divided heterogeneously in both groups: the experimental and control group.

Instruments: Evaluation analysis was developed with the following psychometric tests, operationalized in the variables: “frostig” y “text”:

a) “frostig” variable was measured with the Visual Perception Test of “FROSTIG” (Frostig, 2009).

b) “text” variable was evaluated throughout texts adapted to the curricular competence of each participant.

Variables: The variables: “frostig” and “text” for their three evaluation measures were statistically calculated to form a new variable named “networks” for the three measures: “networks1-2-3” (Table 1).

Table 1: Variables name.

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Procedure: The authorizations and ethical considerations were requested. Then, participants were distributed proportionally in the experimental group and control group.

Over a period of 9 months 3 successive evaluation measures were carried out to both groups: 1) first measure: “frostig1” y “text1” (experimental group began the application of a specific adapted program, while control group continued the regular programs), 2) after 4,5 months, second measurement was made: “frostig2” and “text2, and 3) finally, 4,5 months later, third measure of comparison between experimental group and control group was assessed: “frostig3” y “text3”.

5.6 Data analysis: Results were calculated for the variable: “networks” (1-2-3), in relation to variable “group”, through the ANOVA analysis of repeated measures and independent samples ttest for equality of means, of statistical package SPSS, version 23.

Program

Program was applied to the participants of the experimental group, which get eight general phases [12]:

a. Presentation of global stimulus.
b. Subdivision of initial global stimulus into meaningful parts.
c. Decoding of the previous parts.
d. Concept understanding.
e. Learning links (networks) about the concept.
f. Experiential practices through multisensory processes (psychomotricity, music, video).
g. Perceptual- cognitive reconstruction of the contextual stimulus.
h. Recover information from the learned links.

Results

The sample

A total of 27 participants join in this study, 9 students are diagnosed with ASD1, 9 with ASD2 and 9 with ASD3; 6 students are 4-6 years old, 7 are 7-9 years old, 7 are 10-12 years old, and 7 are 13-15 years; 23 are men and 4 females.

Sample was distributed in two groups: 1 EG (n = 15) and 1 CG (n = 15), that can be seen in Table 2.

Table 2: Participants´ distribution (N = 27).

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Comparative analysis intergroup

In Table 3 can be observed the multivariate test for the 3 successive measurements. Results found significant changes throughout research study (9 months), concluding that there´re meaningful differences for the variable “networks” (value= .92, F= 141.18, error= 24.00, sig= .00). Likewise, the data of 3 measures to the variable “networks” in relation to the variable “group” point that there´re significant differences between participants of the experimental group and participants of the control group (value= .73, F= 34.00, error= 24.00, sig= .00). However, the comparative effects test between groups of Mauchly (Table 4) doesn´t allow refuse the equality hypothesis of means with about the “networks” variable evolution, nor in relationship to the variable “group” (sig = .20), so it´s necessary to complete analysis with the Within-Subjects Effects Tests (Table 5). Indeed, in the Within Subjects Design Test of Table 6, data of previous multivariate analysis are confirmed, in which significant critical levels are found, both, for the variable “networks1-2-3” (sig.= .00, F= 148.15), as well as significant differences in relation to variable “group” (sig= .00, F= 40.38, mean square of error= .205).

Table 3: Multivariate Tests(b).

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Table 4: Mauchly’s Test of Sphericity(b).

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Table 5: Tests of Within-Subjects Effects.

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Table 6: Tests of Within-Subjects Contrasts.

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Data summary of ANOVA is estimated that the evolution of the variable “networks1-2-3” evidence significant differences throughout its evolution (sig= .00, linear F= 242.20) and, also, this evolution is significantly different depending on the participants group type (sig= .00, linear F = 66.86). Indeed, changes found in variable “networks” related to the group type can be seen in Table 7. It looks like experimental group develop from 2.00 in “networks1” to 3.79 in “networks2” and 5.21 in “netswork3” (difference= 3.21 p.). Control group data is slightly lower, since develop from 2.38 in “networks1” to 3.15 in “networks2” and to 3.38 in “networks3” (difference = 1 p.), being total difference between the EG and CG of 2.21 p. The independent samples t- test for equality of means (Table 8) confirms the differences found in the 3 successive measures in the variable “networks” in relation to variable “group”. This test points out that the differences in first measure aren´t significant contrasts (“nextwork1” = .18), second measure also doesn´t find significant different levels: “networks2” (sig = .07), nevertheless, third measure: “networks3”, show significant differences between the experimental group participants and the peers of the control group (sig= .00).

Table 7: Group Statistics.

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Table 8: Independent Samples Test.

Lupinepublishers-openaccess-journals-psychology-behavioral-science

Conclusion

In this research the effectiveness of the applied program could be contrasted, since students of the experimental group have improved significantly, compared to their peers of control group. The improvements found at a comparative level refer to two fundamental aspects:

i. To the levels of sensory integration and gestalt organization, measured through the Frostig test, operationalized in variable “frostig”.
ii. To semantic integration and conceptual understanding processes, evaluated throughout improving the understanding of a text adapted to age and curricular competence, operationalized in variable “text”.

The data found to variable calculated: “nextworks”, which unifies both variables: “frostig” and “text”, allows conclude that the experimental group students have improved sensibly compared to control group peers in sensory perception integration and in processes of semantic coding. Conclusions confirm existence of a neural interrelation between the basic psychological processes during activity of the cognitive development, then specific programs are necessary that facilitate creation of the networks and semantic links, related with the conceptual information and the previously learned concepts. Likewise, it´s essential to forward creation of conceptual categories in semantic memory along the learning process, which globalizes sensory information perceived. Indeed, CG participants also evolved in conceptual creation, but they didn´t significantly as the EG participants, who improved more than their peers of control group, due to application of this program.

Discussion

Indicate that processing of information depends on global perception of stimuli, understanding and semantic analysis, but in people with ASD this process doesn´t occur spontaneously, owing to limited perception and to deficits in construction of interrelated links between the new information and previous contents. Several studies of Cognitive Psychology [13-17] conclude that cognitive functioning is an interrelated systemic process, in which the different basic psychological processes don´t act alone, but that these processes: perception, attention, emotion, motivation and memory interact dynamically during human cognitive action and any component that impedes the functioning of a process, thus, all processes will be affected, not only at the sensory perception, but also in the coding processes. This systemic interaction is developed through intersection of inherent thought process, which is linked to a semantic- pragmatic process about its own functioning dynamics, through which it´s understood and, finally, encoded. In this sense, the analysis and codification of the perceived reality configure a peculiar mode of information processing in people with ASD that influences its global processing mode. There´re previous studies that show the mutual influence of the cognitive process performing how a continuous process that goes from the perception of internal or external stimulus to execution of action itself, therefore, there´s a great number of researches have like aim to ease specific perceptual- cognitive development programs for individuals with ASD [18-22].

Study Limitations

This study is limited by small number of participants because working with people with specific needs is always limited to small samples, so new tests are necessary to corroborate these results.

Acknowledgment

I would like to express my gratitude to the families and professionals of the Association of Families of People with Social Communication Disorders in Ourense, as well as to families and professionals of the educational centers who have participated in this study.

Read More Psychology and Behavioral Science Journal Lupine Publishers Articles:
https://lupine-publishers-psychology-journal.blogspot.com/

Friday 19 February 2021

Lupine Publishers | Cerebral Autosomal Dominant Arteriopathy with Subcortical Infarcts and Leukoencephalopathy (CADASIL) Syndrome: A Case Report

 Lupine Publishers | Journal of Neurology and Brain Disorders


Abstract

Cerebral autosomal dominant arteriopathy with sub cortical infarcts and leukoencephalopathy (CADASIL) is a rare autosomal dominant genetic disease. A35 year old women presented with progressive loss of memory left sided weakness and hemiplegic and non insulin dependent diabetes mellitus with characteristic MRI findings is reported due to its rarity.

Keywords: CADASIL Small vessel disease; Cognitive impairment

Introduction

Cerebral small vessel disease (SVD) is increasingly being recognized as the cause of stroke, impaired cognitive function and mood disorders in geriatric age group. Commonly SVD is sporadic due to old age and high blood pressure, but occasionally SVD has a monogenic cause. The best known among these is cerebral autosomal dominant arteriopathy with sub cortical infarcts and leuko encephalopathy(CADASIL), a rare type of autosomal dominant cerebral angiopathy. It causes recurrent sub cortical ischemic events and vascular dementia which appear as diffuse white matter abnormality on neuro imaging. It involves mainly the small cerebral vessels; histopathology reveals non-atherosclerotic and non-amyloidal angiopathy. [1] The 1st case of aforementioned symptoms was reported 30 years ago in a Swedish family [2] and the term CADASIL was coined in early 1990's [3] Since then CADASIL has been diagnosed in number of families and ethnic groups all over the world.

The prevalence of the condition varies between 2 and 5 in 100,000. Clinically the condition manifests in early or middle adulthood with variety of complaints ranging from migraine, with or without aura, which occur in 30-40% individual, mood disorders, usually depression seen in 30 % people, along with recurrent ischemic events, transient or permanent, accompanied by progressive cognitive impairment leading to dementia and premature death [4-5].

Case Report

This 35 years old married female patient was brought to the psychiatry outpatient department by her family members with the complaint of gradual decline of memory, difficulty in walking due to residual left sided motor weakness and co morbid Non Insulin dependent Diabetes Mellitus. There was no history of hypertension, depression or psychosis or impaired pain sensation. No history of seizures. EEG done earlier did not reveal any abnormality. There was past history of two episodes of CVA in last 5 years. Family history was not significant for headache but there was history of premature death of uncles. Physical examination showed thin built poorly nourished lady. There was no pallor, cyanosis, icterus, edema or generalized lymphadenopathy. Systemic examination of central nervous system revealed grade 3 motor weakness in left upper and lower limbs. Cranial nerves and fundi were normal. There were no sensory deficits or cerebellar signs. Examination of other systems was normal. Score on Mini Mental Status Examination was 12 out of 20. Relevant investigations including hemoglobin, blood sugar, HbA1c, liver function tests, SGOT, SGPT, alkaline phosphatase, blood urea, serum creatinine, were within normal limits. CT Scan brain showed a lacunar infarct in left external capsular region and age inappropriate diffuse cerebral atrophy. MRI of brain revealed diffuse cerebral atrophy with prominent sulcal spaces with periventricular and subcortical arteriosclerotic white matter changes. Small areas of acute to sub acute infarct in rest of the bilateral parietal and frontal lobes support diagnosis of CADASIL Syndrome.

Discussion

CADASIL gene was first identified on the chromosome 19 in the year 1996 [6] Subsequent studies revealed the mutation of NOTCH3 gene on chromosome 19q12. This gene codes for transmembrane receptor protein which is located on surface of smooth muscle cell which surrounds arteries. When pathological NOTCH3 receptor protein got accumulated in cerebral arteries it resulted in varied symptoms of CADAIL [7].

Patient with CADASIL can present with a variety of psychiatric manifestation in 20-41% of cases. The psychiatric manifestations range from agoraphobia, psychoses, personality disorders, alcohol and substance use disorders, adjustment disorder to episodic mood disturbances and bipolar affective disorder [4,8].

CADASIL should be suspected in individual with history of transient ischemic attacks, migraine, severe mood disturbances or late onset affective disorder with headache and neurological symptoms [8-9]. MRI studies shows hyper-intense lesion and ischemic lesion in Basal Ganglia and frontal location of white matter is associated with mood disturbances in patient of CADASIL [10-11].

To conclude it is advisable to do MRI brain in patients presenting with late onset Psychiatric Disturbances and a diagnosis of CADASIL should be considered as a possible differential diagnosis whenever a marked leukoencephalopathy is reported [12].

Read More Lupine Publishers Neurology and Brain Disorders Journal Articles:
https://brain-disorders-lupine-publishers.blogspot.com/

Thursday 18 February 2021

Lupine Publishers | The Dental Service of the French Army in 1940

 Lupine Publishers | Journal of Dental and Oral Health Care


Brief Historical Summary

On December 1 1892, the Brouardel law was voted which gave legal status to dental surgeons. At that point, those who had followed appropriate studies delivered by the Faculty of Medicine were the only ones who could practise dental surgery [1]. Therefore, it was a profession in its infancy which enlisted during the war in 1939? On February 26 1916, a dental service was created in the infantry within the French army for the duration of the war. That of the Navy only started on March 1st 1916. After the Great War, only the Navy decided to dissolve its dental service for it did not have a satisfying organization to greet its patients. In 1934, a backup dental service was implemented once again. While they were made lieutenants in 1818, dentists could become captains after the law voted on December 19 1934 [2]. A few months later, dentists in the land army were attributed the same rank.

National Federation of Reserve Dental Surgeons (Federation Nationale Des Chirurgiens-Dentistes De Reserve - Fncdr)

In 1925, the first dental officers who returned to civil life assembled to form the Association of dentists of the land army, the navy and the air force from the Paris region during the 3rd Congress of medicine and military pharmacy. It was aiming at keeping the ties they built during the war and at helping each other to get back to normal life [3]. A two-year optional superior military training was created. Once students in dental surgery had passed it, it allowed them to accomplish the very last part of their military service as 2nd class military dentist which was a rank similar to that of second lieutenant. Then, they joined dental care garrisons, centres for edentulous patients or even centres for maxillofacial prosthodontics. In 1927, there were only eight of them. Those who had not passed their two-year optional superior military training were only able to join the nursing unit [4,5].

Thereafter, these associations asked the Minister of War and of National Defence the permission to open an advanced level training school for dental surgeons called EPOR. The central administration of the French Defence Health service (Direction centrale du Service de santé des armées -DCSSA), who was in charge of diversifying the health service’s branches, immediately supported the initiative which led to the opening of the first EPOR. This training school offered its first services during the monthly lectures of the Villemin military hospital on October 1926. From 1931, the initiative taken by Paris spread over the country and each military region could benefit from their own EPOR. Moreover, integrating this school soon became a prerequisite to get an upper grade [6].

In 1933, all those regional associations eventually gathered into a National federation of Reserve Dental surgeons which was notably later chaired by Pierre Budin who was one of the eleven chairmen of the federation and who received the Legion of Honour in 1920 for his bravery on the battlefield. This institution thoroughly devoted itself to the improvement of the dentists’ status. The DCSSA organized the dental surgeon’s training in order to get them ready for all sorts of duties: anaesthetists, surgical assistants, radiologists, first-aid workers, etc. It was Pierre Henry’s case who was mobilised in 1940 and who joined the surgical team of the Ancemont Hospital where he practised as anaesthetist. During the war, as he was demobilised, he practised in Rennes [7]. It was in this way that, in the Val-de-Grâce Hospital, the general practitioner Ginestet trained dental surgeons to specialties of maxillofacial surgery. The dentists were informed on various medical practices with respect to the damage caused by more modern weapons and which led to more complex and numerous maxillofacial injuries. All that work which was organised with the help of the DCSSA prepared the intervention of the dental surgeons’ corps in the event of future conflicts [4,5].

Active Dentists

The French Defence Health service which included the dental surgery service reshaped its organisation following the assessments which were carried out at the end of World War I. It decided to implement more mobile structures which aimed at following a regiment or a division or which were able to adapt to the new schemas of so-called movement wars. Therefore, mobile surgical groups, advanced hospitals, medical hospitals of evacuation and various health services appeared. Within each of these structures, one or two dentists were incorporated. Through Article 39, the law of April 1 1923 planned that dental students had to accomplish their military service in the health service as nurses if they had not finished their dental curriculum or as second class dental surgeons if they had graduated [7].

Mobilisation

World War II broke out on September 3 1939. Everybody mobilised in France. Reserve dental surgeons were called upon as lieutenants for the most of them on different strategic areas. They were either assigned in the sanitary formation of the divisions, in the surgical ambulances or in the front regiments. At the beginning of the war, they barely practised dentistry. They mainly served as assistants for specialised medical treatments. Only a few practised in the garrison dental surgeries or in those of hospitals. Indeed, with the growing mobilisation, the Health service created dental surgeries in each area settlements and in the most important garrisons. Thus, in the front, they intervened in aid stations to cure the wounded. In divisional areas, they collaborated with doctors and surgeons within mobile or advanced surgical groups. Finally, at the rear, they were employed by maxillofacial centres or by radiology services. At the beginning of the war, immediate interventions and first-aid treatment had priority and demanded a lot of efforts especially as the bombing victims were not only soldiers but also civilians. Those who had followed the teaching delivered by the EPOR were more easily assigned in surgical services [6].

(Delivered) Treatment

Curing the soldiers’ oral pathologies, soothing their pain and quickly making them operational was a primal objective. A man who is in pain does not fight right and immediately constitutes a threat for his comrades. Furthermore, a person with bad teething cannot chew properly which can lead to other general diseases. Seeing the soldiers and being among the troops also allowed the dental surgeons to deliver notions of oral hygiene. A schematic note was written for each soldier in order to keep track of his treatment and the head of the dental surgery also used to write reports on a daily basis. On them were written all the consultations and carried out interventions. Reconditioning oral cavities was systematically carried out before each military operation with a compulsory preliminary tartar removal which was the first hygienic approach. Numerous dental extractions were carried out. Each attribution of a prosthetic device had to be subjected to the approval of the head of the Regional Health Service through the main stomatologist’s intermediary.

Once the head of the dental surgery gave his approval the extractions were made before the making of a device. A month and a half later, the device could be made. On the contrary, if the making was refused, only extractions were made [6]. In each dental surgery, the staffs was made up of a department head who often was a stomatologist or a military dental surgeon and of military dental surgeons to assist him according to the needs decided by the main regional stomatologist. This department head was under the supervision of the chief doctor of the training to which he was linked such as the corps troops, the regimental infirmary, the military hospital, etc., while being under the main stomatologist’s command [7]. In 1940, continuing education was insured thanks to magazines such as “Odontologie” which ended its publication between June/ July 1944 and March/April 1945. In 1942, the publications spread for lack of promoting advertisement and because of the war, and eventually diminished until they did not get enough cover.

Stomatology Service

During World War II and after the French surrender which was signed in Rethondes on June 22 1940, numerous demobilised dental surgeons were called upon to treat soldiers who were wounded in the face. This was when maxillofacial surgery became extremely important in terms of medical care. It became a full-fledged specialty which benefited from new techniques and materials. Stomatology services existed along the secondary organisations such as the garrison dental surgeries which operated in each area settlements and in the most important garrisons. They were implemented at the beginning of the war.

There were two different articulations concerning the rear structures:

Inter-Regional Technical Services

The inter-regional centre of surgery and of maxillofacial prosthetics was independent from the regional service but was under the supervision of the chairman of the regional health service which it was a part of. A surgeon, head of department, was at its head and benefited from a stomatologist’s assistance who was leading the maxillofacial prosthetics service. As a team, they could rely on all the specialists and dental prosthetics technicians. In these inter-regional centres, they mainly treated damaged maxillary bones as well as those of the face and the neck thanks to this coeducational surgical and stomatological training. On the contrary, simpler maxillary fractures, that is to say those without a loss of substance, were treated in the regional stomatological centres. Likewise, simple edentulous patients were not accepted in surgical and stomatological centres. Centres for edentulous patients were implemented for the wounded that could be fitted with prosthesis. Each centre was specialised. The main objective was to quickly orientate the wounded.

Regional Technical Services

Only one stomatology service existed in a region. In the event of a big region, there could be more of them. Only one centre of dentures for edentulous patients worked to making the dentures of all the edentulous people who came in the dental surgeries of the local garrison. Those local stomatology and dental prosthetic services contained from 50 to 100 beds out of a total of around 10,000 and 20,000 beds for the entire medical sector. The odontostomatologic illnesses too important to be treated in the dental surgeries, the dental illnesses, simple maxillary or prosthetic fractures, and surgical extractions were operated there. The simple fitting of prosthesis on edentulous people was also carried out there. The management of regional centres was insured by the main local stomatologist. He had to keep a close watch on the stomatology and dental prosthetic centre, and dental surgeries. Here was his staff:

i. A dental prosthetist who was able to make a set of dentures per day, that is to say 30 dentures per month;

ii. A military dentist who was able to provide the same work than that of 4 out of 5 technicians.

Figure 1: Dental surgery in Fort Hackenberg-Maginot line (20 km away from Thionville, Mozelle).

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 2: Dental surgery in Four à Chaux, Vosges.

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 3:

Lupinepublishers-openaccess-dentistry-oral-healthcare

An officer was in charge of the management of the equipment which included the maintenance and the stock according to the wants and needs. He also kept a log of which was fitted in the patients ‘mouths with the description of the device. This allowed justifying the price of each set of dentures. Finally, the local head could subsequently follow the activity of the centre for edentulous patients thanks to a monthly report [8]. The ability training of the military service of December 7 1938 for metropolitan troops included a calculation giving the soldier’s chewing coefficient. In case of a lack of needed dentures, the fitting was made possible only if the soldier’s chewing coefficient was lower to 25%. Therefore, each fully edentulous man could receive a full set of dentures and could be kept in the service (Figures 1-3).

In case of maxillofacial emergency, the wounded soldier had to be driven to a maxillofacial centre from two to ten days after the injury. The quicker the treatment started the better were the chances of recovery. As for the repairing phase, it often needed transplants. The immobilisation of fractured bony edges was essential to be successful. The average period of the primary consolidation of transplants was of around two months but it was only around the fourth month that the transplants looked normal again. On November 11 1942, the German army invaded the free zone, France’s Southern part. A month later, Vichy’s “Armistice Army” was disbanded. Hence, the French Defence Health service no longer existed. Any medical support was entrusted to the British and mostly to the Americans during the conflict. In 1940, numerous demobilised dentists put up resistance, their dental surgeries often served as turntables for the exchange of information. Many of them were shot by the Gestapo [9]. Many others were deported and came back weakened from the camps (René Maheu)(Figure 4). Some others enlisted in the Free French Forces (Maurice Prochasson). Dental lieutenant’s uniform (purple velvet on the flat-topped French military cap, on the collar tips, the caduceus was visible on each collar tips and buttons), made between 1930 and 1935, worn in 1940, the khaki military cap was made from 1930. Before it was blue (Figure 5).

Figure 4:

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 5:

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 6:

Lupinepublishers-openaccess-dentistry-oral-healthcare

The braids were on the turn-ups during World War I. They were above the turn-ups from 1934/1935. Here, on the picture, they are on the turn-ups. The dental officer’s uniform was similar to that of the cavalry (executive order 1934)(Figures 6 & 7). Here, it was prior to 1934. Among the decorations, from left to right, the following ribbons are recognisable:

Figure 7:

Lupinepublishers-openaccess-dentistry-oral-healthcare

i. Knight of the Legion of Honour.

ii. 14-18 Cross of War (without mention in despatches).

iii. Combatant Cross (created following the law of June 28 1930).

Figure 8: Pierre Audigé (1908-1944) (Audigé Family, 2005).

Lupinepublishers-openaccess-dentistry-oral-healthcare

iv. Impossible to identify, but given the order of wearing official French decorations, it could be the Allied Medal (or Victory Medal) created following the law of July 20 1922 and which was granted provided three successive months or not of presence between August 2 1914 and November 11 1918, to the soldiers belonging to the units enumerated by the ministerial instruction of October 7 1922, which is worn before distinction (Figures 8-11).

Figure 9: Alain Maheu (1899-1980) back from deportation (Maheu, 2003).

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 10: Maurice Prochasson (Museum of the Order of the Liberation, 2011).

Lupinepublishers-openaccess-dentistry-oral-healthcare

Figure 11: A soldier wounded in the face upon his arrival (left) and his departure (right) of the African centre (Converse J., Péri M. & Roche G. K. T., 1944).

Lupinepublishers-openaccess-dentistry-oral-healthcare

v. Médaille commémorative française de la Grande Guerre

vi. The Insignia for the Military Wounded. The trousers have a command band which existed from 1830 [10-13].

Read More Lupine Publishers Dentistry and Oral Health Care Articles:
https://lupine-dentistry-oral-health-care.blogspot.com/

Saturday 13 February 2021

Lupine Publishers | Synthesis and Characterization of Poly (1,4-Benzenedimethylene Phthalate) and the Study of its ability to sorb Pb(II), Cd(II), and Zn(II) ions

 Lupine Publishers | Journal of Organic and Inorganic Chemical Sciences

Abstract

Poly (1,4-benzenedimethylene phthalate) was synthesized by condensation of phthaloyl chloride with 1,4-benzenedimethanol in the presence of pyridine in dry THF at 30 °C . The resulting polymer white powder was characterized by viscosity measurement, FT-IR, 1H and 13C NMR, elemental analysis, thermal (TGA-DSC) methods and scanning electron microscopy. The uptake properties of Pb(II), Zn(II) and Cd(II) metals by the polymer from aqueous solutions were studied by the batch and column techniques as a function of pH, temperature, concentration and contact time. The uptake increased with increasing pH reaching a maximum at pH = 6.00 for all ions. The polymer showed high uptake capacities toward Pb(II) and Zn(II) ions, but medium uptake capacity toward Cd (II). The linearized forms of Langmuir, Freundlich and Dubinin-Radushkevich adsorption isotherms indicate studied ions Pb(II), Zn(II), and Cd(II). The adsorption capacity follow the order: Pb(II) > Zn(II) > Cd(II).

The thermodynamic functions, ΔGǂ, ΔHǂ and ΔS° were determined for Pb(II), Zn(II) and Cd(II); the values of AG° indicated that the adsorption process of these metal ions on the polymer is favorable. In this work, the values of ΔGǂ, ΔGǂ and ΔSǂ were determined by using Eyring-Polanyi equation and Arrhenius equation. The results indicated that ΔGǂ=72.8, 78.6, 87.8kJ/mol for Pb, Zn, Cd ions, respectively. The column experiments for metal ion uptake were conducted at pH= 6.0, 25.0 °C, and initial concentration of 150.0mg/L. The loaded concentrations were 76.42, 49,05 and 28.47 ppm for Pb, Zn, Cd ions, respectively. The efficiency for recovery of metal ions after adsorption by treatment of the loaded polymer with 0.1M HNO3, 0.1M EDTA, gave good percent recovery for 0.1M HNO3

Keywords: Benzendimethanol; Phthalate polymers; Condensation polymerization; Polymers uptake; Metal ion uptake; Adsorption

Introduction

Heavy metals are known for their toxicity to living organisms. Their concentrations in the environment increase with the increase of industrial activities. Most important, these metals are not biodegradable and tend to accumulate in living organisms, and causes serious health problems [1]. Allot of methods were reported for removal of metal ions from aqueous solutions; these include filtration, adsorption, chemical precipitation and ion exchange [2]. Different approach for removal of metal ions from water streams which have been increased in recent years [3], is the adsorption of metals by polymers. This approach is characterized by high effectiveness, high adsorption capacity, and high selectivity, fast kinetics of metal-ion uptake, high mechanical strength, chemical resistance and recyclability of the polymer [4,5].

According to their methods of preparation, synthetic polymers can be divided into three major types:

    A. Chain Reaction (Addition) Polymers: These polymers are formed by addition to the double bond of monomers containing carbon-carbon double bonds, called vinyl monomers, without the loss of small molecules. The repeating units of addition polymers have the same composition as the monomer [6].

    B. Condensation Polymers: these polymers are formed by condensation reaction of monomers containing organic functional groups. These reactions usually involve elimination of small molecules (e.g., water, methanol, and ethanol) [6].

    C. Other Polymers May Be Prepared by Modification of Polymers by: I- Attachment of an appropriate specific ligand groups to the polymer chains. This method has been preferred since functional groups containing ligands can easily be attached to polymers than inorganic supports. [7,8]f The attachment of dithiocarbamates to polystyrene cross linked with 2% divinylbenzene is an example of this type [9]. II- Polymerization of monomers to which ligand groups are attached. Polymers with ester linkages in their main chain are used in many applications such as biomedical matrices, liquid crystals, fibers, and heat resistant materials [10]. Polyesters are characterized by higher adsorption capacities, efficiencies as well as high selectivity to some metal ions. [11] Synthetic polyesters may be prepared by polycondensation of diacids with diols, diaciddichlorides with diols and ester interchange reaction of diesters with diols.

There are two types of polyesters: Aliphatic polyesters and aromatic polyesters. The synthesis of aliphatic polyesters has been well established for several years. However, these polyesters possess low thermal stability due to their low melting points and glass transition temperatures owing to their low molecular weight. These properties resulted in limited usage and few applications of aliphatic polyesters, yet they showed potential as biodegradable polymers. On the other hand, aromatic polyesters display an excellent pattern of physical properties. They are strongly resistant to hydrolysis, bacterial and fungal attack, they also remain unaltered in the environment [11,12], Combining aromatic and aliphatic units in the same polyester chain has been envisaged as an attractive approach to obtain novel products encompassing biodegradability and high performance properties [11].

There are a lot of polymers were prepared and their sorption behavior was studied in our laboratory,such aspoly (1,4-cyclohexanedimethylene oxalate), Poly (bisphenol A oxalate), poly-cis,trans (1,3-cyclohexylene oxalate), poly (bisphenol- Aphthalate), poly (bisphenol-Acis-1,2 cyclohexanedicarboxylate), poly (1,4-yclohexanedimethanolphthalate) and poly (1,4-cyclohexanedimethanol succinate).

Experiment

Materials

The chemicals were obtained from commercial sources as either analytical reagent grade or chemically pure grade and were used as received. The chemicals were purchased from the corresponding companies: Phthaloyl dichloride, 1,4-benzenedimethanol, cadmium(II) nitrate tetrahydrate and zinc(II) nitratehexahydrate from BDH; N,N-4-dimethylaminopyridine (4- DMAP), dimethylsulfoxide (DMSO) from Acros, pyridine (GPR), nitric acid (65%), hydrochloric acid (36.5 %), and disodium ethylenediaminetetraacetate(EDTA)(Scharlau), sodiumperchlorate (SIGMA), tetrahydofuran (THF) (GCC) and lead(II) nitrate (PRS Paureac) and 4-dimethylaminopyridine (4-DMAP)from Acros.

Preparation of The Polymer

Poly(1,4-benzenedimethylenephthalate) was synthesized by polycondensation using single phase organic solvent polymerization. 1,4-benzenedimethanol ( 4.97g, 0.036 mol), pyridine (8.54g, 0.108mol) and a catalytic amount of 4-DMAP were dissolved in THF (60 mL). To this solution, a solution of phthaloyl chloride (7.307 g, 0.036mol) in THF (30mL) was added drop wise with stirring. The reaction mixture was stirred for 1 h at (30-35) oC and then for 3 days at room temperature. The polymer precipitated as a white solid from the THF solution. The solvent was evaporated and the solid was dissolved in chloroform (150mL) and washed with water (2x500mL), (6% v/v) HCl solution (1 x 150mL), and finally with distilled water (3x500mL). The chloroform solution was dried over anhydrous sodium sulfate, and was then concentrated to about 100 mL of solution. The polymer was precipitated by drop wise addition of chloroform solution to 500.0mL of methanol. The precipitated polymer was then filtered and dried at 55.0 oC under vacuum to give a white powder 66.0g, (68 % yield).

Preparation of Stock Solutions

Stock solutions (1000.0 mg/L) of the three metal ions were prepared by dissolving specific amounts of the salts of Pb(II), Zn(II), and Cd(II), in 0.1 M NaClO4 which was adjusted to the desired pH. The stock solutions were used to prepare solution with different concentrations (20.0, 40.0, 50.0, 60.0, 80.0, 100 and 150.0mg/L). The dilution is achieved by using 0.10 M NaClO4 and adjusted by 0.10 M HClO4 to achieve the desired pH= 4.00, 5.00 and 6.00.

Study of Metal Uptake Characteristics of The Polymers By Batch Technique

The metal uptake characteristics for each metal ion were studied using batch equilibrium technique. An aqueous solution of known metal ion concentration (25.0mL) was added to polymer powder (0.10 g), the mixed solutions were mechanically shaken, after a certain period of time at 25.0 oC, 35.0 oC, 45.0 oC, the mixture was filtered and the amount of the metal ion remaining in the filtrate solution was determined by atomic absorption spectrometry after constructing up an analytical calibration curve for each element (Pb(II), Zn(II), and Cd(II)).

The Rate of Metal Ion Uptake

Experiments for determining the equilibrium time for the metal ion uptake process involving 100.0mg ± 0.1 mg of the polymer was swelled with 25.0mL of metal ion solution containing 150mg/L metal ion at different pH, and the solution was mechanically shaken. The contact time was varied from 5 minutes to 48 hours at 25.0 oC, 35 oC, 45 oC. The mixture was filtered and the amount of the metal ion remaining was determined with atomic absorption spectrometry. The amount of metal ion uptake by the polymer (qe), may be obtained from the following relation:

Lupinepublishers-openaccessjournals-chemical-science

qe: Metal ion uptake by the polymer (in mg M(II)/ g polymer).

Ci: Initial metal ion concentration (mg/L).

Ceq : The residual concentration of the metal ion in solution at equilibrium (in mg/L).

V: volume of solution (L).

m: mass of polymer (g).

And the percentage of metal ion loading by the polymer expressed as % uptake was [13]:

Effect of pH on The Metal-Ion Uptake

Similar experiments were also carried out, under different pH values of 4.00, 5.00 and 6.00 for fixed contact time of 24 hours to determine the effect of pH on the metal ion uptake by the polymer.

Adsorption Isotherms Studies

The adsorption of Pb(II), Zn(II), and Cd(II) was carried out by taking a known mass of 100.0g ± 0.1mg of the polymer swelled with 25.0mL of solutions of concentration variation ranging from (20.0- 150.0)mg/L for each metal, under different pH values of 4.00, 5.00 and 6.00 and different temperatures (25.0, 35.0 and 45.0) oC.

Metal Ion-Uptake By The Polymer Using Column Experiment

Glass column of 30.0 cm length and 1.5cm inner diameter was used in this experiment. The column was packed with 1.00g ± 0.1mg dried polymer. A sample volume of 150.0mL containing Pb(II) of 1000mg/L was passed through the column at a flow rate of 1.0mL/4min. The eluate was collected in a 100.0mL volumetric flask, and concentration of the metal ion was then determined by AAS. The same experimental conditions were used for the determination of Zn(II), and Cd(II) ions uptake, where the sample which passed through the column was 150mg/L of these metal ions.

Desorption studies

The desorption of the Pb(II), Zn(II), and Cd(II), ions was carried under column condition, where the polymer was loaded with each metal ion as described before, using 50.0mL of two eluting agents, 0.10M HNO3 and 0.10M EDTA were used for polymer recovery from adsorbed metal ion, keeping the flow rate of elution at (1mL/4min). The concentration of metal ion in the eluate was collected in five 10.0mL portions, and was then determined by AAS.

Lupinepublishers-openaccessjournals-chemical-science

Water regain (α):

Water regain is defined as the amount of water absorbed by 1000.0mg ± 0.1 mg of polymer [14]. A sample of dry polymer was suspended in water, and was left for 2 and 24 hours. The polymer was filtered and weighed, dried at 60.0 oC and then re-weighed. Water regain (α) was calculated from the mass difference (eq).

Lupinepublishers-openaccessjournals-chemical-science

Results And Discussion

Polymer synthesis

The synthesis of poly (1,4-benzenedimethylene phthalate), from equimolar amounts of 1,4-benzenedimethanol and phthaloyl chloride was performed by solution polycondensation in THF at 30.0 °C in the presence of excess pyridine as acid scavenger and 4-DMAP as the catalyst. The reaction proceeded by pyridine- catalyzed nucleophilic displacement of the chloride of the phthaloyl chloride with the alcoholic group of 1,4-benzenedimethanol. The relatively high yield of the polymer may have been due to the high reactivity of phthaloyl chloride group. The resulted polymer, which was obtained as a powder was found to be insoluble in many common organic solvents such as tetrahydrofuran (THF), diethyl ether, acetone and methanol but soluble in chloroform.

Solution viscosity

The inherent viscosity of poly(1,4-benzenedimethylene phthalate) solution was calculated from viscosity measurements of dilute polymer solutions (0.5g/dL) in chloroform at 25 °C. The polymer had an inherent viscosity of 0.22dL/g. This value indicates that the polymer had an intermediate inherent viscosity which implies that it had moderate molecular masses. This value is higher than the those obtained at 0-5° for poly(bisphenol-Aphthalate), poly(bisphenol-Asuccinate), and poly(cyclohexanedimethylene phthalate) which had the values of 0.11dL/g , 0.13dL/g, and 0.11dL/g, respectively [15].

Infrared spectroscopy

The polymer was analyzed by FT-IR spectroscopy. The FT-IR spectrum exhibits characteristic absorption bands for the major bonds involved in the polymer. The FTIR spectrum (Figure 1) showed two strong absorption bands for the stretching vibration of the carbonyl group (C=O) of the phthalate ester group at 1726cm-1 and for C-O-C in the range from 1124 to 1279cm-1. Another strong IR band was observed at 2951cm-1 assigned to the C-H stretching in the 1,4-benzendimethanol moiety. These wave numbers, which are typical for the ester group are conformed to the reported literature [15], and thus confirm the formation of the postulated polymer Scheme 1 and Figure 1.

Scheme 1: Preparation of poly (1,4-benzenedimethylene phthalate) polymer.

Lupinepublishers-openaccessjournals-chemical-science

Figure 1: IR spectrum of poly(1,4-benzendimethayl phthalate) polymer.

Lupinepublishers-openaccessjournals-chemical-science

The Nuclear magnetic resonance (nmr) spectra for polyesters

Figure 2: 1H-NMR spectrum of the polymer.

Lupinepublishers-openaccessjournals-chemical-science

1H-NMR spectrum: The polymer was analyzed by NMR spectroscopy in order to elucidate its chemical structure and support its formation. In the 1H-NMR spectrum of poly(1,4- benzendimethylene phthalate), the signal of the aromatic protons of the phenylene ring of benzendimethylene unit was observed as singlet at δ = 7.36ppm, whereas that of the aliphatic methylene protons attached to the oxygen of the ester group was observed as singlet at δ = 5.22ppm. The aromatic methylene protons of the phthalate unit in ortho and meta positions to the ester were observed at δ = 7.71ppm and 7.50ppm, respectively. The 1H-NMR spectrum is shown in Figure 2. In the 13C-NMR spectrum of poly(1,4- benzendimethylene phthalate), the signal of the aromatic carbon atoms of the phenylene ring of benzendimethylene unit appeared at δ = 129ppm. The signal of the quaternary aromatic carbon atoms appeared at δ = 136ppm. The signal of the aliphatic methylene carbon atoms attached to the oxygen of the ester group appeared at δ = 67ppm. The signal of the quaternary aromatic carbon atoms of the phthalate unit to which the ester group is attached appeared at δ = 132ppm. The signals of the aromatic carbon atoms in ortho and meta positions to the ester group appeared at δ = 129ppm and 133ppm, respectively. The signal of the carbonyl carbon atom of the ester group appeared at δ = 167 ppm. The 13C-NMR spectrum of the polymer is shown in Figure (2), and the1H-NMR and 13C-NMR spectral data for the polymer assigned to the various proton and carbon atoms are presented in Table 1 (Figures 3 & 4).

Figure 3: 13C-NMR spectrum of the polymer.

Lupinepublishers-openaccessjournals-chemical-science

Figure 4: DSC thermo gram of polymer.

Lupinepublishers-openaccessjournals-chemical-science

Table 1: Thermal stability of polymer.

Lupinepublishers-openaccess-journals-chemistry

Thermal Properties

The thermal properties of polyesters synthesized were also investigated by differential scanning calorimetry (DSC), and thermogravimetric analysis (TGA) under dry N2 atmosphere.

Thermal Transition

The thermal properties of polymer were investigated with DSC and TGA. The Tg value of the polymer was 52°C. This value is considered lower than expected for such an aromatic polymer such as poly(ether carbonate) containing aromatic -aromatic ether showed Tg values from room temperature up to 47°C [16] and polyquinoxalines and other aromatic polymers were studied Tg from 215.5 to 394.5°C [17]. This Tg value may have been due to the imparted flexibility effects of the aliphatic methylene groups of the 1,4 -benzenedimethanol. The presence of the aliphatic moieties in the polymer backbone imparted flexibility to polymer segments to move under the effect of temperature. This ease of motion is reflected in the slightly low Tg value of the polymer [16]. The DSC thermogram of the polymer is shown in Figure 4.

Thermal stability

The thermal stability of the polymer was investigated by TGA under dry nitrogen. Table 1 summarizes the initial thermal decomposition (onset) temperature Tid, T5%d , T10%d , and T50%d decomposition temperatures, which correspond to the temperatures at which 1, 5, 10, and 50% loss of mass of polymer occurred, respectively. The table 2 also shows the residual mass percent remaining after heating the polymer to 673.1 °C which was found to be 8.74%. The thermogram of the polymer Figure 5 displayed a typical one-stage characteristic with a relatively fast mass loss occurring at temperatures between 400 and 450 °C. The fast mass losses may have been due to decomposition of the polymer backbone. These values are higher than the corresponding values obtained in the case of poly (1,4-cyclohexandimethlene- phthalate), which occurred between 350-450°C and the residual mass equal 0.44% [15].

Table 2: Several reported metal uptake values by chelating polymers in the literature.

Lupinepublishers-openaccess-journals-chemistry

Figure 5: TGA thermogram of poly(1,4-benzenedimethayl phthalate) polymer.

Lupinepublishers-openaccessjournals-chemical-science

Morphological Characterization

The surface ofthe polymer was also characterized by SEM before and after metal sorption. The SEM micrographs for the polymer synthesized are shown in Figure 6. The pores are distributed on the rough surface of the polymer, the small pores which are similar to the forms of flowers are located on the surface of the rods with an average size <5|im. The existence ofthese pores provides convenient diffusion channels for metal ions into the interior of the polymer when it is used for adsorption of metal ions from aqueous solution. The interior structure of the polymer showed randomly distributed large gaps and air pockets created during polymerization. The spongy structure of the inner rods maximize the contact surface between the polymer and the solution which led to increased metal ion uptake, this is shown in Figure 6a for the polymer before metal sorption. However, after adsorption of metal ion by the polymer, a slight loss occurred to the composition of surface features and of the channel and a small part of its surface became smooth. This is represented for the polymer surface loaded with Cd(II) ions and is shown in Figure 6b & c. It has been found by SEM investigation that loading the polymer with zinc ions changed its surface topography, increased the proportion of smooth surface and resulted in narrow channels and pores Figure 6d & 6e. The highest uptake of metal ions by polymer was for Pb(II), the polymer loaded with lead ions was studied by SEM to observe the changes in surface topography that took place, the entire surface became smooth and the pores disappeared due to full metal ion coverage. SEM for our polymer was similar to that of poly(cychlohexandimethelen succinate) (Al- Dweri, 2010). The images are presented in Figure 6f-6i.

Figure 6: SEM images of poly(1,4-benzenedimethelen phthalate).

Lupinepublishers-openaccessjournals-chemical-science

Water Regain or Water Content ( α )

Water regain experiment was performed to determine the water regain ratio (α) for the polymer. Water regain is usually correlated with the hydrophilic character of the polymer, the higher the water regains, and the more hydrophilic the polymer is. The water regain value for the polymer was found to be 0.019 g/g after 2h of stirring and 0.029g/g after 24h stirring, this indicates that the polymer has a low hydrophilic character. These values are higher than the corresponding values obtained by poly(1,3-cyclohexylene oxalate) polymer which was found to have values of 0.011g/g and 0.014g/g, respectively. Based on these values the polymer is considered to be low hydrophilic in nature [18] and these values are smaller than the corresponding values obtained for poly (1,4-cyclohexanedimethylene oxalate) polymer which was found to have values of 0.064g/g and 0.087g/g [19]. Principally, the water molecules are polar and would interact with the polar groups of the polymer (the ester group), this interaction explains the water regain properties of the polymer.

The rate of metal ion uptake by the Polymer

The adsorption kinetics of metal ions on the surface of the polymer was investigated as shown in Figure 7 for example. The adsorption of metal ions increase with time until complete saturation.

Figure 7: Metal uptake as a function of contact time, at pH = 6, T = 45 °C, and initial concentration of 150ppm.

Lupinepublishers-openaccessjournals-chemical-science

Adsorption Isotherms of the Polymer

Using linearized Langmuir and linearized Freundlich isotherm as analyzed to determine adsorption isotherms for metal ion Pb(II), Zn(II) and Cd(II) at different pH values (4.00, 5.00 and 6.00) and different temperatures (25.0 oC, 35.0 oC and 45.0 oC) in the range of concentrations from 20.0 to 150.0mg/L. The adsorption isotherms results are presented in Table 3 and shown in Figure 8 & (Table 2) (20-23).

Figure 8: Plots of a) adsorption isotherm b) Linearized Freundlich c) Linearized Langmuir of Pb(II) at pH = 6 and 45 °C.

Lupinepublishers-openaccessjournals-chemical-science

Table 3: R2, qm, and KL values obtained from Langmuir plots

Lupinepublishers-openaccess-journals-chemistry

In the Langmuir model, the values of correlation regression coefficient (R2) are greater than 0.90 and had excellent linearity. This indicates that homogenous sites of interaction are better to describe the process and the maximum sorption capacities (qm) deduced from these results indicated that the polymers has the highest capacity towards Pb(II) ions, but it shows a lower capacity towards Cd(II). It is observed that the adsorption capacity (qm) increases as the temperature and pH values increase for all investigated metal ions. Comparing the values of qm of the synthesized polymer to literature values Table 2. Typical qm values obtained in this work at pH= 6.00 and 45.0 oC for the adsorption of various metal ions on our polymer were as follows:

Metal ion: Pb(II) > Zn(II) > Cd(II)

qm (mg/g) = 35.7 > 23.8 > 22.7

Metal ion: Zn(II) > Cd(II) > Pb(II)

qm (mmol/g) = 0.36 > 0.20 > 0.17 (Tables 3 & 4).

Table 4: R2, KF and n values obtained from Freundlich plots

Lupinepublishers-openaccess-journals-chemistry

The trend in qm values in poly(1,4-cyclohexanedimethylen phthalate) at pH= 6.00 and 25.0 oC is Pb(II) > Cd(II), in (mg/g) were 53.5>17.7 respectively [15]. The trend in qm values for poly(1,4- cyclohexanedimethylene oxalate) at pH= 4.00 and 25.0 oC is Pb(II) > Cd(II) > Zn(II), in (mg/g) were 31.2 > 29.8 > 15.9, respectively [19]. The trend for Pb(II) and Cd(II) ions in poly(hydroquinone oxalate), and in poly(neopentyl oxalate) polymer, were as follows: Pb(II)>Cd(II) [24-27] and thus the results similar in order the metal ion with our results in the literature [23] that is qm values 207.7 >30.7 >19.1 for Pb(II) >Zn(II) >Cd(II) in poly(bisphenol A oxalate). The results indicate that our polymer has reasonable qm values. In the Freundlich model, both KF and nare Freundlich constants, being indicative of the adsorption capacity and the adsorption intensity respectively. High value of n between (1.4-3.6) indicates that adsorption is good over the entire range of concentration studied, while small values of n means that the adsorption is good at high concentrations but much less at lower concentrations and the values of n were greater than one indicating that the adsorption was favorable. A greater value of KF indicates a higher capacity for the adsorption than smaller values [28].

Dubinin-Radushkevich

The linear form for Dubinin-Radushkevich (D-R) isotherm has the following expression:

Lupinepublishers-openaccessjournals-chemical-science

R is the ideal gas constant (8.3145 J.mol-1.K-1) and T is the

absolute temperature (Kelvin). The values of p and q max are evaluated from the slope and intercept of the linear plot of lnq versus s2, where q max is related to the adsorption capacity and p is the constant related with the adsorption energy. The free energy of adsorption (E) is defined as the free energy change required for transferring one mole of ions from solution to the solid surface, this energy is calculated as follows:

Lupinepublishers-openaccessjournals-chemical-science

Table 5: Dubinnin-Radushkevich (D-R) isotherm parameters for Pb(II), Zn(II) and Cd(II).

Lupinepublishers-openaccess-journals-chemistry

As illustrated in Tables 3-5, the values of E for Pb (II) are (0.1100.600 kJ/mol), for Cd(II) are (0.080 -0.200 kJ/mol) and for Zn(II) are (0.180-0.310kJ/mol). All values are less than 8.00kJ/mol, this indicates that physical forces affect adsorption. The result of the concentration variation isotherms of polymer (Table 5) and the plots of Dubinin-Radushkevich at pH= 4, 5 and 6 at 25 °C, 35 °C, 45 °C are shown in Figure 9.

The Effect of Temperature on The Uptake

The effect of varying temperature on the % uptake of metal ions was also investigated. The results are obtained by plotting % uptake at different pH against temperature and are presented in Figure 10. The results obtained showed that the adsorption process of Pb(II) , Zn(II) and Cd(II) onto the surface of the polymer is an endothermic process since the % uptake increases as the temperature increases nearly at all tested pH values.

Figure 9: Plots of Dubinnin-Radushkevich (D-R) isotherm for the adsorption of Pb(II) on the polymer at pH= 6.00 and a) T= 25 °C, b) T= 35oC, c) T=45 °C

Lupinepublishers-openaccessjournals-chemical-science

Arrhenius Parameter

Table 6: The Arrhenius parameters for Pb(II), Zn(II) and Cd(II) at pH=6 and temperatures 25.0oC, 35.0oC, 45.0oC.

Lupinepublishers-openaccess-journals-chemistry

Table 7: The Arrhenius parameters: Activation energy ( E) and pre-exponential factor (A) for Pb(II), Zn(II) and Cd(II) at pH=6 and temperatures 25oC, 35oC, 45oC.

Lupinepublishers-openaccess-journals-chemistry

Arrhenius equation can be used to determine the activation energy and pre-exponential factor for a reaction were calculated from Figures 11 and shown in Table 6. The results A and Ea values that obtained in Figure 11 and Table 6 are calculated from Arrhenius equation:

Ln k = - Ea / R T + Ln A .....(4) (Table 7)

Where k is the rate constant, Ea activation energy, R gas constant and A pre-exponential factor. The values of A and Ea can be calculated from intercept and slope of a straight line of a plot of Ln k against 1/ T. The pre-exponential factor A is the constant of proportionality between the concentration of the reactants and the rate at which they collide. The activation energy Ea is the minimum kinetic energy required for a collision to result in reaction, through the more favorable molecular orientations. The factor exp (-Ea / RT) represents the fraction of molecular collisions that have an energy value equal to or greater than the activation energy Ea. At higher temperatures a larger portion of reactant molecules will have the required Ea to react. Thus, the reaction rates depend on Ea, the reactant orientations (relative positions) during collisions and the temperature. Both A and Ea are approximately constant over a moderate range of temperature (50K) [29]. The order of Ea values for Pb, Zn and Cd is (19.2, 32.1, 83.9kJ/mol) respectively. This low value of Ea indicates a reaction rate slightly sensitive with temperature [30]. So, we cannot explain the mechanism of the process from activation energy only but needed to use Eyring equation [31].

The Eyring equation (activated complex theory)

Determination of activation energy, entropy and enthalpy of activation by this equation

Lupinepublishers-openaccessjournals-chemical-science

Thus, the above general form ofthe Eyring equation or activated Complex Theory equation, also known as Eyring-Polanyi equation in chemical kinetics, relates the reaction rate to temperature and is trivially equivalent to the empirical Arrhenius equation. The results were calculated from Figures 11 & 12 and shown in Table 8.

Figure 10: The temperature dependence of % uptake of Pb(II) , Zn(II) and Cd(II) ions at pH=6.

Lupinepublishers-openaccessjournals-chemical-science

Figure 11: The Arrhenius Plots of (a) Pb(II), (b) Zn(II) and (c) Cd(II) at pH= 6.00.

Lupinepublishers-openaccessjournals-chemical-science

Figure 12: The Activated Complex theory Plots of (a) Pb(II), (b) Zn(II) and (c) Cd (II) at pH= 6.00 .

Lupinepublishers-openaccessjournals-chemical-science

Table 8: Calculated thermodynamic parameters of activation (AH, ASt, AGt) at pH= 6 and 25oC.

Lupinepublishers-openaccess-journals-chemistry

The Eyring-Polanyi equation has been applied to rate processes and the calculation of values of enthalpies and entropies of activation without pointing out the significance of the obtained values and the difference between the activation energy values found by using the Arrhenius equation.

Lupinepublishers-openaccessjournals-chemical-science

A plot of (Ln k/T) versus 1/T gives a straight line with a slope of -ΔHǂ/R from which the enthalpy of activation can be derived and with intercept of ln (kB/h) +ΔSǂ/R from which the entropy of activation is derived. From the values of the free energy of activation the real energy requirements are known, thus, suggesting the use of the Eyring-Polanyi equation mainly as a tool for gaining a deeper understanding of the actual processes at work and not only as a tool for predicting reaction rates based on measured rate constants. This relation is usually used for the suggestion of a mechanism for a certain reaction in the following way: the certain reaction is performed at various temperatures where the reaction rate constant is measured. The pre-exponential factor A of Arrhenius equation has been related to ΔSǂ of Eyring equation.

Low values of lnA correspond to negative values of ΔSǂ, the activated complex in the transition state has a more organized, more ordered and more rigid structure than the reactants. This happens when bonds are formed or substances are absorbed, and high values of lnA correspond to positive (or less negative) values of ΔSǂ, a positive value for the entropy of activation indicates that the transition state is disordered (less organized), compared to the state of the reactants. This happens when bonds are broken or substances are desorbed. The calculated value of the entropy of activation is used for the suggestion of a mechanism i.e. in replacement reactions: Associative (ΔSǂ< 0), Dissociative (ΔSǂ> 0), Interchange (ΔSǂ = 0). [32] In our results in Table 8, the values of ΔSǂ for Pb(II), Zn(II) and Cd(II) are -188.7, -164.7, -21.3 J/mol.K, respectively all the values of ΔSǂ< 0 indicated that the process has Associative mechanism. The relation between Ea of the Arrhenius and ΔHǂ of the Eyring equation (activated complex theory) is:

Lupinepublishers-openaccessjournals-chemical-science

There was no comparison between Ea values and the ΔGǂ values that lead to the conclusion that Ea does not represent the full energetics of a process. Thus ΔGǂis the critical factor and not Ea. Thus, near room temperature (the thermodynamic temperature, 25 oC), Ea is roughly 2.5 kJ mol-1 larger than ΔHǂ. [32] Through Table (4.68) can be calculated divide between ΔHǂ and Ea for Pb(II), Zn(II) and Cd(II) that equal (2.6, 2.6, 2.5 kJ mol-1) respectively, there results indicated of the temperature independence on reaction rate and the temperature influences is room temperature 25 oC [32]. For, so, we needed calculate the free energies of activation ΔGǂ. It has been found that ΔGǂ gives a more realistic/true value of the "activation the processes need in order to take place and not Ea or ΔSǂ alone. The free energy of activation ΔGǂ includes not only the ΔHǂ component (= Ea-RT) but also the ΔSǂ component that may be important. The term -TΔSǂ that hΔSǂo be added to ΔHǂ in order to give ΔGǂ which may be critical,

ΔGǂ = ΔHǂ- T ΔSǂ......(8)

Determines the spontaneity of the reaction ΔGǂ

ΔGǂ greater than zero reaction is spontaneous

ΔGǂ0 = system at equilibrium, no net change occurs

ΔGǂ less than zero reaction is not spontaneous [33]. The results of ΔGǂ in Table 8 were explaining that all the values of ΔGǂ > 0. So, this process is spontaneous (physical process).

Distribution Coefficient (Kd)

The distribution coefficient is defined ΔSǂhe final concentration of metal ion in the sorbed form on polymer divided by its final concentration in solution. It is regarded a standard parameter in the assessment of the physicochemical behavior of metal ions between solid and liquid phases. It is calculated by the following equation.

Kd = qe / Ce = KL qm-KL qe........(9)

Table 9: Distribution coefficient for Pb(II) for different pH values and temperatures.

Lupinepublishers-openaccess-journals-chemistry

Where Kd is the distribution coefficient (L/g). Thus, a plot of (qe / Ce) against (qe) should be a straight line with slope= - K and an intercept= qm K if Langmuir equation is applicable The distribution coefficients (Kd) is calculated for the polymer at different pH values (4.0, 5.0, and 6.0) and temperatures (25°C, 35°C, and 45°C) are given in Table 9.

Thermodynamics of Adsorption on the Polymer

In order to understand the possible adsorption mechanism involved in the removal process, thermodynamic functions for the system, including changes in Gibbs free energy (ΔG°), change in enthalpy of adsorption (ΔHǂ) and changes in entropy of adsorption (ΔS°), were calculated using the following equation Using the following equation:

Lupinepublishers-openaccessjournals-chemical-science

Where Kd is the equilibrium constant, R is the gas constant and T is the temperature in Kelvin.

The plot of lnKd against 1/T for each metal ion gives a linear relationship, where the values of enthalpy (ΔHo) were entropy (ΔSo) are obtained from the slope and intercept of lnKd vs. 1/T plots. AGo was calculated at each temperature using the following equation:

Lupinepublishers-openaccessjournals-chemical-science

Table 9 present the data of distribution coefficients at different temperatures and pH values, while Figures 13 show the plots of lnKd versus 1/T (Table 10).

Figure 13: Plots of ln KdVs 1/T for Pb(II),(a) pH =4.00 (b) pH=5.00 and (c) pH=6.00.

Lupinepublishers-openaccessjournals-chemical-science

Table 10: Thermodynamic Functions for Pb(II) at T= 298.15 (K).

Lupinepublishers-openaccess-journals-chemistry

The results of the studies on the influence of temperature on metal ions adsorption are presented in Table 10 above. The positive values of enthalpy indicate that the adsorption removal increased with increase of the solution temperature. This shows that the adsorption process is an endothermic one. AG°, ΔHǂ, and ΔS° are the thermodynamic functions related to the experiment conditions. Spontaneity and favorability of the adsorption process is established by decrease in Gibbs free energy values, AG°. The value of AG° between (0.16 - 1.04) indicates the degree of favorability of the adsorption process, so the values of AG presented in Table 10 indicate that the adsorption of Pb(II), Zn(II) and Cd(II) is a favorable process [28]. All the values of AG° are very small and positive which suggests that the adsorption of metal ions onto polymer require some small amount of energy to convert reactants into products [34].

This is agreeing with values of Table 10 which represent to the degree of favorability of adsorption, the decrease of AG° values of Pb(II) >Zn(II) >Cd(II). As shown in Table 10 all AH values are positive this suggests the endothermic nature of metal adsorptions. One possible explanation of this is the well-known fact that heavy metal ions used are well solvated in water. In order for these ions to be adsorbed, they are denuded to some extent of the hydration sheath. This dehydration process of ions requires energy for removal of water from ions is essentially an endothermic process [35]. We assume that the energy of dehydration exceeds the exothermicity of the ions attaching to the surface [36]. The implicit assumption here is that after adsorption the environment of the metal ions is less aqueous than it was in the solution state.

The endothermic interactions between polymer surface and metal ions were accompanied by small positive values of entropy, which wΔS⫲he driving force for adsorption. The positive values of AS signify an increased state of randomness at the solid- solution interface following adsorption. Also the positive entropy of adsorption reflects the affinity of adsorbent for metal ions used. The adsorbed water molecules, which are displaced by the adsorbate species, gain more translational energy than is lost by the adsorbate ions, thus allowing the prevalence of randomness in the system [28]. The entropy changes were most likely to be due to structural changes and adjustments in the adsorbate as well the adsorbent. The structural changes arise from the release of ions like H+ from the polymer surface into the solution and also from partial solvation of the metal ions in water. The adsorptions of Pb(II), Cd(II) and Zn(II) on polymer were associated with entropy decrease in conformity with the general situation of ions existing in a more chaotic random distribution in aqueous solution compared to their adsorbed and immobilized states [37].

Column Experiments

Metal ion uptake by the polymer

The metal ion uptake by the polymer using column experiment for Pb(II), Zn(II) and Cd(II) was determined at pH 6.0 and 25.0 oC, initial concentration of 150.0 mg/L and a flow rate 1 mL/4min. The uptake for metal ions is represented in Table 11.

Table 11: Metal iron uptake using column experiments.

Lupinepublishers-openaccess-journals-chemistry

It can be seen that the uptake capacities of the polymer with the metal ions fall in the order;

Lupinepublishers-openaccessjournals-chemical-science

This result is similar to the order of the metal ions in the batch experiment. However, the values of percent uptake for the metal ions in column experiment are lower than those obtained in batch experiments, because in order to achieve the complete saturation a much greater time is needed. On the other hand, there is no mechanical shaking associated with the column experiments, which result a decrease in percent of metal uptake.

Desorption Studies

Two eluting agents, 0.10 M HNO3 and 0.10 M EDTA were used for removal of metal ions, keeping the flow rate of elution 1 mL/4 min. The fluent was collected in five portions, 10.0mL for each portion; the results are expressed as percent recovery and represented in Table 12. The eluting agents react in two different ways: HNO3 act as proton-exchange agent and the second a complex-forming agent as EDTA. Depending on the values of the % accumulative recovery, in Table 12, the following trend was observed for the eluting agents of metal ions from the polymer:

0.1M HNO3 > 0.1M EDTA

Table 12: Desorption of Pb(II), Zn(II) and Cd(II) ions from the polymer.

Lupinepublishers-openaccess-journals-chemistry

The experiments confirmed that maximum metal desorption can be achieved with mineral acids in concentrations of 0.1M solutions. This could be attributed to cation exchange between the proton and metal sorbed. However, this method is more complex than protonation [8,38].

Conclusion

In this study, we prepared a polymer containing phthalate function group and capable of adsorbing the metal ion by solution polycondensation and characterization of poly(1,4- benzenedimethelene Phthalate). The structure and properties of polymer was confirmed by FT-IR, 1H NMR, 13C NMR, elemental analysis, SEM and thermal analysis. The sorption properties of the synthesized polymer toward Pb(II), Zn(II), and Cd(II) in aqueous solutions were examined under various experimental conditions using both batch and column experiments. The effective desorption for the metal ions was studied, and the coefficient of recovery of sorption ability was also investigated. The polymer has high sorption rate for Pb(II) observed during the first 24h with high percentage of uptakes toward Pb(II), Zn(II) and low percentage of uptakes toward Cd(II) ions. The influence of different pH on metals uptake showed that the metal-ion uptake by the polymer increased with increasing pH and reached a maximum at pH=6 for Pb(II), Zn(II), and Cd(II). The best conditions for adsorption of metal ions and maximum adsorption capacity (qm) on polymer surface are pH=6, T= 45°C and initial metal concentration of 150 ppm.

The obtained adsorption data showed fitting for Langmuir, Freundlich and Dubinin-Radushkevich adsorption isotherm models. The application of the Eyring equation to literature data i.e. the calculation of AH+, AS+ and AG+, has pointed out that in geochemical transformations it is necessary to calculate the entropy of activation along with the enthalpy of activation in order to fully characterize a process energetically. A column packed with the polymer has good metal uptake properties toward all metal ions, and followed the order: Pb(II) > Zn(II)> Cd(II) at pH 4.0 and 25 oC and flow rate 1 mL/4min. The efficiency of recovery of metal ions after adsorption can be carried out by treatment of the loaded polymer with 0.1M HNO3 and 0.1M EDTA with good percent recovery.

Acknowledgements

The authors would like to thank the Deanship of Academic Research and Quality Control (DAR) of The University of Jordan for supporting this work and all supervisors who worked in the improvement of this work.

Read More Lupine Publishers Chemistry Journal Articles:
https://lupinepublishers-chemicalsciences.blogspot.com/